Measured parameters

  • ECG (3/5-lead)

  • RESP (respiration rate)

  • Pulse rate

  • NIBP (non-invasive blood pressure)

  • SpO₂ (pulse oximetry)

  • TEMP (temperature) – optional

  • EtCO₂ (end-tidal CO₂) – optional

  • Advanced SpO2 anti-interference.
  • Detect and Analyze up to 28 types of arrhythmia events.
  • Low perfusion SpO2 algorithm with visual alarm.
  • Multi-Lead ECG algorithm with ST & QT analysis.
  • Unibody cable – one cable for SPO2, ECG and Temperature
  • Ergonomic & Concise Design.
  • EWS Early Warning Score: 
  • * AIView VX patient monitor provides a point-of-care EWS (Early Warning Score) tool to assist clinicians or caregivers in recognizing the early signs of clinical deterioration to initiate early intervention.
  • * Compliant with National Early Warning Score 2 (NEWS2), Modified Early Warning Score (MEWS), and Cardiac Arrest Risk Triage Score (CART).
  • * Initiative display of scoring value and radar-gram on the screen to remind clinicians anytime.
  • Excellent and Valuable IT Solutions.
  • Flexible Mounting and Carrying Solutions.

Detailed parameters : 
ECG
Leads: 5-electrode leads: I, II, III, aVR, aVL, aVF, V
Waveform sweeping speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Sampling rate: 500 Hz

CMRR: =90 dB
Differential input impedance: =10 MO
Input signal range: -10.0 mV to +10.0 mV (peak-to-peak value)
HR measuring range:
adult: 15-300 bpm, ±1 bpm whichever is greater
pediatric: 15-350 bpm, ±1 bpm whichever is greater
neonate: 15-350 bpm, ±1 bpm whichever is greater

SpO2
SpO2 measurement range: 0-100% ± 2%

Wavelength: red light: 660 nm, infrared light: 905 nm
Maximum optical output power: =2 mW
Perfusion index (PI): 0.1%- 20%

PR
Display range: 30-250 bpm
Measurement range from NIBP: 30-250 bpm, ±3 bpm or ±3%, whichever is greater
Measurement range from SpO2: 30-250 bpm, ±2 bpm or ±2%, whichever is greater

Temp
Measurement range: 0.0-50.0°C; ±0.1°C Temp
Measurement method: skin temperature
Measurement time: <150s

NIBP
Maximum single measurement: adult/pediatric <180 s; neonate <90 s
Static measurement range: 0-300 mmHg; ±3 mmHg
Initial inflating pressure:
adult: 80 mmHg-280 mmHg
pediatric: 80 mmHg-210 mmHg
NIBP measuring range:
SYS: 25-290 (adult) 25-240 (pediatric)
DIA: 10-250 (adult) 10-200 (pediatric)
MAP: 15-260 (adult) 15-215 (pediatric)

Resp
Measuring range: adult: 0-120 rpm; pediatric/neonate: 0-150 rpm
Waveform gain: ×0.125, ×0.25, ×0.5, ×1, ×2, ×4
Waveform speed: 50 mm/s, 25 mm/s, 12.5 mm/s, 6.25 mm/s

 

Product Type Patient monitor
Display 10-inch touchscreen, detachable screen
Measured Parameters ECG, RESP, pulse rate, NIBP, SpO₂
Optional Parameters Temperature, EtCO₂
Connectivity / Analysis Arrhythmia, ST and QT analysis listed in supplied data